Approval for the ovation abdominal stent graft system. this device is indicated for treatment of patients with abdominalaortic aneurysms having vascular morphology suitable for endovascular repair, including: 1) adequate iliac/femoral access compatible with vascular access techniques, devices, and/or accessories; 2) non-aneurysmal proximal aortic neck: a) with a length of at least 7 mm proximal to the aneurysm; b) with an inner wall diameter of no less than 16 mm and no greater than 30 mm; and c) with an aortic angle of <= 60 degrees if proximal neck is >= 10 mm and <=45 degrees if proximal neck is <10 mm. 3) adequate distal iliac landing zone: a) with a length of at least 10 mm; and with an inner wall diameter of no less than 8 mm and no greater than 20 mm.