PubMed:1010249 JSONTXT 4 Projects

[Malfunction of the Smeloff-Cutter mitral prosthesis: echo-phonocardiographic diagnosis in two successfully re-operated cases (author's transl)]. La malfunzione della protesi mitralica di Smeloff-Cutter. Diagnosi eco-fonocardiografica in due casi rioperati con successo Five cases of Smeloff-Cutter prosthesis implanted in the mitral position nine years ago were investigated with combined echocardiographic and phonocardiographic recordings. In two cases a malfunction was demonstrated and a surgical confirmation was obtained at the reoperation. Normal and abnormal echocardiographic patterns of the Smeloff-Cutter prosthesis are reported and discussed. The Authors stress the importance of combined echo-phono recordings when malfunction of the mobile element of a mechanical prosthesis is suspected.

Annnotations TAB TSV DIC JSON TextAE

  • Denotations: 0