PubMed:29212681 / 1024-2094 JSONTXT 3 Projects

Annnotations TAB TSV DIC JSON TextAE

Id Subject Object Predicate Lexical cue
T3 0-1070 RESULTS denotes There was a significantly reduced rate of steady-state linear wear with XLPE (0.07 mm/yr) compared with UHMWPE (0.37 mm/yr) (p = 0.001). Volumetric wear was also significantly reduced in the XLPE group (29.29 mm3/yr) compared with the UHMWPE group (100.75mm3/yr) (p = 0.0001). There were six patients with UHMWPE who had non-progressive osteolysis and none in the XLPE group. All three bearing groups had significant improvements in 12-item short form health survey scores, Western Ontario and McMaster Universities Osteoarthritis Index score and Harris Hip Score. However, the improvement in HSS was significantly less in the UHMWPE group (p = 0.0188) than in the other two groups. At ten years, the rates of volumetric and linear wear in the XLPE group remain low and predominantly below the estimated threshold for osteolysis (1 mm/yr). The rate of linear wear in the XLPE group was three times less than in the UHMWPE group at five-year follow-up and five times less at ten years. The rate of volumetric wear was also three times less in the XLPE group at ten years.