T285 0-107 Sentence denotes With its good sensitivity for samples up to CT ≈ 30, the colorimetric RT-LAMP assay has several advantages:
T286 108-179 Sentence denotes It is fast, inexpensive, and it can be evaluated without any equipment.
T287 180-428 Sentence denotes RT-LAMP reactions also appear to be less sensitive to contaminants in the samples than RT-qPCR, but care has to be taken that the samples used do not alter the pH as the colorimetric RT-LAMP assay is performed under conditions of weak pH buffering.
T288 429-593 Sentence denotes Some clinical samples contain contaminants that can lead to acidification of the reaction independent of the presence of a template RNA if too much sample is added.
T289 594-871 Sentence denotes Diagnostic RT-qPCR tests usually include a technical internal control, i.e., another RNA species, which is spiked into all samples and which is detected independent of the gene of interest to safeguard against the possibility of a general reaction failure within a sample tube.
T290 872-945 Sentence denotes It would be desirable to have a similar precaution for the RT-LAMP assay.
T291 946-1074 Sentence denotes A multiplexed fluorescence readout might provide this (34) but comes at the expense of the simplicity of a colorimetric readout.