T131 176-375 Sentence denotes RT-qPCR–positive samples with a CT < 30 scored positive in the RT-LAMP assay (79 of 81), whereas almost all samples with CT values between 30 and 40 scored negative (only 4 positive of 36) (Fig. 3B).
T132 376-498 Sentence denotes This confirmed the sensitivity of the RT-LAMP assay for detection of SARS-CoV-2 RNA in samples corresponding to a CT < 30.
T133 499-657 Sentence denotes We observed small differences between different plates on the exact sensitivity threshold, probably caused by slight variability in plate or reagent handling.
T134 658-821 Sentence denotes We found two RT-qPCR–negative samples that scored positive in the RT-LAMP assay (Fig. 3A and Table 1) and one sample that scored just below the ΔOD cutoff of +0.3.
T135 822-910 Sentence denotes The overall specificity of the RT-LAMP test was 99.7% (Wilson’s 95% confidence interval:
T136 911-1027 Sentence denotes 98.9 to 99.9%), and the sensitivity for samples with CT < 30 on RT-qPCR was 97.5% (Wilson’s 95% confidence interval:
T137 1028-1066 Sentence denotes 91.4 to 99.3%) (Fig. 3B and table S2).