PMC:7574920 / 17164-20486 JSONTXT 7 Projects

Annnotations TAB TSV DIC JSON TextAE

Id Subject Object Predicate Lexical cue
T108 0-73 Sentence denotes Validation of the colorimetric RT-LAMP assay for SARS-CoV-2 RNA detection
T109 74-182 Sentence denotes To determine the specificity and sensitivity of the RT-LAMP assay, we continued to analyze more RNA samples.
T110 183-258 Sentence denotes We assayed a total of 768 RNA samples obtained on different days (fig. S1).
T111 259-545 Sentence denotes Visualization of the RT-LAMP assay results 30 min after the start of the incubation at 65°C showed comparable behavior of the samples in a total of ten 96-well test plates (Fig. 3A and Table 1), indicating that the RT-LAMP assay was reproducible from day to day and from plate to plate.
T112 546-606 Sentence denotes Fig. 3 Detection of SARS-CoV-2 RNA using the RT-LAMP assay.
T113 607-732 Sentence denotes (A) Scatter plot shows a comparison of RT-LAMP assay results and RT-qPCR results for RNA samples tested on 10 96-well plates.
T114 733-847 Sentence denotes The RNA extraction method (QC, QiaCube, a column-based method; QS, QiaSymphony, a bead-based method) is indicated.
T115 848-963 Sentence denotes The time point for measurement by the colorimetric RT-LAMP assay was 30 min after the start of the 65°C incubation.
T116 964-1019 Sentence denotes The 96-well plate shown in Fig. 2 is not included here.
T117 1020-1245 Sentence denotes Table 1 shows numbers of samples stratified according to the results of the RT-LAMP and the RT-qPCR assays. (B) Sensitivity (right) and specificity (left) of the RT-LAMP assay [derived from data in (A) and Table 1] are shown.
T118 1246-1360 Sentence denotes The specificity is the fraction of RT-qPCR–negative samples correctly identified as negative by the RT-LAMP assay.
T119 1361-1654 Sentence denotes For sensitivity, the RT-qPCR–positive samples were stratified by CT values into three bins (as indicated by x axis labels), and for each bin, the sensitivity is given as the fraction of qPCR-positive samples in the respective CT bin that have also given a positive result in the RT-LAMP assay.
T120 1655-1896 Sentence denotes The thick black lines indicate the values of these fractions (i.e., the specificity and sensitivity estimates); the black boxes indicate the corresponding 95% confidence intervals (Wilson’s binomial confidence interval). (See also table S2).
T121 1897-2011 Sentence denotes Table 1 Shown is RT-qPCR and RT-LAMP testing of 768 clinical samples stratified into CT value bins (see Fig. 3A).
T122 2012-2103 Sentence denotes Fig. 3B and table S2 show specificity and sensitivity values calculated from these numbers.
T123 2104-2111 Sentence denotes RT-LAMP
T124 2112-2131 Sentence denotes CT Pos Neg Sum
T125 2132-2163 Sentence denotes RT-qPCR Pos 0–25 51 0 51
T126 2164-2180 Sentence denotes 25–30 28 2 30
T127 2181-2197 Sentence denotes 30–35 4 16 20
T128 2198-2214 Sentence denotes 35–40 0 16 16
T129 2215-2237 Sentence denotes Neg Neg 2 649 651
T130 2238-2431 Sentence denotes Sum 85 683 768 The consistency of the results during the analysis confirmed a threshold of ΔOD > +0.3 as a robust measure to identify samples that were positive for SARS-CoV-2 RNA (Fig. 3A).
T131 2432-2631 Sentence denotes RT-qPCR–positive samples with a CT < 30 scored positive in the RT-LAMP assay (79 of 81), whereas almost all samples with CT values between 30 and 40 scored negative (only 4 positive of 36) (Fig. 3B).
T132 2632-2754 Sentence denotes This confirmed the sensitivity of the RT-LAMP assay for detection of SARS-CoV-2 RNA in samples corresponding to a CT < 30.
T133 2755-2913 Sentence denotes We observed small differences between different plates on the exact sensitivity threshold, probably caused by slight variability in plate or reagent handling.
T134 2914-3077 Sentence denotes We found two RT-qPCR–negative samples that scored positive in the RT-LAMP assay (Fig. 3A and Table 1) and one sample that scored just below the ΔOD cutoff of +0.3.
T135 3078-3166 Sentence denotes The overall specificity of the RT-LAMP test was 99.7% (Wilson’s 95% confidence interval:
T136 3167-3283 Sentence denotes 98.9 to 99.9%), and the sensitivity for samples with CT < 30 on RT-qPCR was 97.5% (Wilson’s 95% confidence interval:
T137 3284-3322 Sentence denotes 91.4 to 99.3%) (Fig. 3B and table S2).