T300 0-188 Sentence denotes First, to determine the temperature window of operation, MDA reactions are performed at 25 °C and 30 °C using the Illustra GenomiPhi V2 DNA amplification kit and EvaGreen fluorescence dye.
T301 189-338 Sentence denotes From the literature, we know that this reaction does not work above 35 °C due to degradation of the protein activity in presence of a substrate [44].
T302 339-485 Sentence denotes In Figure 13, a graph of the fluorescence signal during MDA reactions at 25 °C and 30 °C, together with their non template control (NTC) is shown.
T303 486-861 Sentence denotes These reactions are carried out in a conventional Bio-Rad CFX96 Touch Real-Time PCR machine (Bio-Rad Laboratories, Inc., Hercules, CA, USA) and the results show that the chosen proof-of-principle DNA amplification reaction is temperature dependent to some extent, but that there is a wide range of temperatures at which the amplification can be performed, i.e., 25°C to 35°C.
T304 862-977 Sentence denotes This makes the functioning of the integrated resistive heater less critical than the stability shown in Figure 11c.