PMC:7143804 / 36594-38487 JSONTXT 7 Projects

Annnotations TAB TSV DIC JSON TextAE Lectin_function

Id Subject Object Predicate Lexical cue
T283 0-4 Sentence denotes 3.4.
T284 5-21 Sentence denotes Chip Functioning
T285 22-162 Sentence denotes Characterization of the actual heat distribution is done using a FLIR One Pro iOS thermal camera (FLIR Systems, Inc., Wilsonville, OR, USA).
T286 163-325 Sentence denotes Thermal images of the heat distribution are made at the side of the substrate without the resistor, whereas different input powers are used to heat up the heater.
T287 326-532 Sentence denotes Au reflects the infrared radiation of the environment directly, therefore an image with the resistor facing the camera would give a heat map of the surrounding and not of the real temperature of the heater.
T288 533-633 Sentence denotes These measurements also gives a better insight of the heat distribution inside the reaction chamber.
T289 634-698 Sentence denotes The images are processed using the FLIR postprocessing freeware.
T290 699-755 Sentence denotes Results of these measurements are shown in Figure 11a,b.
T291 756-822 Sentence denotes The results are in good agreement with the estimations in Table 1.
T292 823-1205 Sentence denotes The slight deviation between the values can be explained by the fact that the heated area in the calculations had an assumed value, the thermal camera measurements used 1.5 mm thick COC substrates without a water-filled chamber, the actual resistors have small parts wich have a double thickness due to the two used shadow masks, and rounding of the values used in the calculations.
T293 1206-1329 Sentence denotes The reliability of the heater is tested by inserting the thermocouple into the temperature control chamber (see Figure 3a).
T294 1330-1442 Sentence denotes A constant input potential of 4 V is applied using the Keithley source and the temperature is measured for 25 h.
T295 1443-1594 Sentence denotes This exceeds the required operation time at least twelve-fold, meaning that it is a good indication for the reliability of the heater and thermocouple.
T296 1595-1631 Sentence denotes The results are shown in Figure 11c.
T297 1632-1893 Sentence denotes To perform on-chip amplifications, the resistive heater on the chip is connected to the Keithley source using crocodile connections and the thermocouple is inserted in the temperature control chamber and connected to a Tenma 72-7715 Thermometer (see Figure 12).