Id |
Subject |
Object |
Predicate |
Lexical cue |
T75 |
0-197 |
Sentence |
denotes |
For evaluation of assay performance regarding efficiency, linearity and unspecific signals, we used the SuperScript protocol in a direct comparison of 73 samples with the RealStar SARS-CoV-2 assay. |
T76 |
198-430 |
Sentence |
denotes |
The E gene assays showed 97% (71/73) identical results in both assays (60 negative and 11 positive results as well as two divergent results, both with a positive Real Star SARS-CoV-2 assay and a negative SuperScript protocol assay). |
T77 |
431-556 |
Sentence |
denotes |
The specific SARS-CoV-2 assays gave in 67 (92%) identical results (60 negative and seven positive) and six divergent results. |
T78 |
557-622 |
Sentence |
denotes |
Overall, the RealStar SARS-CoV-2 kit seemed to be more sensitive. |