T63 0-544 Sentence denotes Three additional one-step real-time PCR systems were compared with the initially used QuantiTect Virus +Rox Vial kit (QIAGEN) (Table): (i) OneStep RT-PCR kit (QIAGEN; data not shown in the Table because only a limited number of samples were tested in only one run), (ii) LightCycler Multiplex RNA Virus Master (Roche, Mannheim, Germany; data not shown because only a limited number of samples were tested in only one run) and (iii) SuperScript III One-Step RT-PCR system with Platinum TaqDNA Polymerase (Invitrogen, Darmstadt, Germany) (Table).
T64 545-707 Sentence denotes Each assay protocol and thermoprofile was adjusted to the recommendations of the manufacturer, but primer and probe concentrations were the same as published [5].
T65 708-785 Sentence denotes They all showed comparable results in reduction of unspecific E gene signals.
T66 786-1147 Sentence denotes Using SuperScript III, the unspecific signals in the E gene screening assay were significantly reduced to 5% (13/257) in tested patient samples and negative extraction and non-template controls (Table), whereas Ct values for positive controls (Wuhan coronavirus 2019 E gene and SARS-CoV Frankfurt 1 genomic RNA [3]) were in the same range in QuantiTect (median:
T67 1148-1160 Sentence denotes 29.1; range:
T68 1161-1207 Sentence denotes 26.8–32.4; n = 85) and Superscript (median Ct:
T69 1208-1220 Sentence denotes 28.1; range:
T70 1221-1247 Sentence denotes 26.4–31.0; n = 30) setups.