PubMed:12477021 JSONTXT

Annnotations TAB JSON ListView MergeView

{"target":"https://pubannotation.org/docs/sourcedb/PubMed/sourceid/12477021","sourcedb":"PubMed","sourceid":"12477021","source_url":"https://www.ncbi.nlm.nih.gov/pubmed/12477021","text":"Clinical performance of the Reichert AT550: a new non-contact tonometer.\nThe aim of the present study was to assess the level of accuracy for measurements of intra-ocular pressure (IOP) obtained with a new non-contact tonometer (NCT) the Reichert AT550. Measurements were compared against those obtained with the Reichert Xpert Plus, Goldmann applanation tonometer and Perkins tonometer. Thirty-five university students were assessed with the four tonometers in a randomised order, with non-contact tonometry performed first. Each of the four measurement devices had its own trained clinical observer. Plots of differences of IOP as a function of the mean for each pair of instruments were obtained. No statistically significant differences were found when comparing the AT550 NCT with contact applanation tonometry (AT) (p\u003e 0.05), displaying the closest level of agreement (as represented by the lowest mean difference and the narrowest confidence interval) with the Goldmann tonometer (limits of agreement, 0.12+/-2.17). In conclusion, readings of IOP with the AT550 NCT are clinically comparable with those obtained with Goldmann tonometry in a population with IOP within the normal range.","tracks":[]}