[Comparison of three medical goggle sterilizing approaches]. |
| T2 |
61-71 |
Sentence |
denotes |
OBJECTIVE: |
| T3 |
72-138 |
Sentence |
denotes |
To compare three sterilizing methods for reusable medical goggles. |
| T4 |
139-147 |
Sentence |
denotes |
METHODS: |
| T5 |
148-252 |
Sentence |
denotes |
A total of 180 medical goggles of the same brand and same model were randomly divided into three groups. |
| T6 |
253-578 |
Sentence |
denotes |
In group A the goggles were first soaked with 2000 mg/L chlorine-containing disinfectant and then cleaned manually; goggles in other two groups were sterilized using pre-programmed automatic spray cleaning and disinfection machine, the disinfection program was set to 90 ℃ for 5 min in group B and 70 ℃ for 30 min in group C. |
| T7 |
579-715 |
Sentence |
denotes |
The quality of the sterilization was monitored by visual inspection with luminous magnifying glass and residual protein detection assay. |
| T8 |
716-871 |
Sentence |
denotes |
User satisfaction on cleanliness of medical goggles, clarity of mirror surface and suitability of elastic bands was investigated with questionnaire survey. |
| T9 |
872-880 |
Sentence |
denotes |
RESULTS: |
| T10 |
881-1102 |
Sentence |
denotes |
The qualification rates verified by visual inspection were 82.4%, 84.6%and 98.3%in group A, B and C, respectively, the qualification rate in group C was significantly higher than those in group B and group C (all P<0.05). |
| T11 |
1103-1243 |
Sentence |
denotes |
The qualification rates verified by residual protein detection assay were 96.7%, 100.0%and 100.0%in group A, B and C, respectively (P>0.05). |
| T12 |
1244-1303 |
Sentence |
denotes |
A total of 54 questionnaires were submitted for the survey. |
| T13 |
1304-1464 |
Sentence |
denotes |
The satisfaction rates were 100.0%, 90.7%and 94.4% for cleanliness of medical goggles, clarity of mirror surface and suitability of elastic bands, respectively. |
| T14 |
1465-1477 |
Sentence |
denotes |
CONCLUSIONS: |
| T15 |
1478-1595 |
Sentence |
denotes |
Machinery sterilization set 70 ℃ for 30 min has better cleaning and sterilizing effects for reusable medical goggles. |