{"target":"http://pubannotation.org/docs/sourcedb/PMC/sourceid/8950092","sourcedb":"PMC","sourceid":"8950092","source_url":"https://www.ncbi.nlm.nih.gov/pmc/8950092","text":"The results from tests using configuration ‘c’, with an air curtain, are shown in Figure 14. The first test had the highest phage concentration at Sampler F, 4347 PFU/L Air. Sampler F consistently collected high virus numbers of during all three tests. Figure 14 shows the average particle flow based on the three tests conducted using this configuration, with the highest average phage number collected at Sampler F, 5600 PFU/L Air. For this configuration, particles were concentrated at the patient’s head. The p-values obtained from ANOVA analysis suggested that there was no significant statistical difference between the three tests performed.","tracks":[]}