{"target":"http://pubannotation.org/docs/sourcedb/PMC/sourceid/7532132","sourcedb":"PMC","sourceid":"7532132","source_url":"https://www.ncbi.nlm.nih.gov/pmc/7532132","text":"FTIR analysis was conducted using a Perkin Elmer Frontier FTIR spectrometer to identify any interactions occurring between molecules, such as hydrogen bonding. Diamond ATR crystal was wiped clean with ethanol, and a background scan was carried out. The sample was then placed onto the crystal. The pressure was applied at a force gauge of 70. A total of 16 scans were taken at a resolution of 2 cm−1 and using a wavelength range of 4000–600 cm−1. Two replicates were carried out for each sample.","tracks":[]}