The high-resolution microscope detection system (Optique Peter, France) was based on a 50μm thick LuAG:Ce scintillator imaged with a 10× magnifying microscope objective onto a sCMOS sensor (pco.edge 5.5, PCO, Germany), resulting in an effective pixel size of 0.65μm. |
T296 |
337-497 |
Sentence |
denotes |
The high photon flux density allowed for image acquisition with continuous motor movement, short acquisition time of 35 ms per frame, and a framerate of 20 fps. |
T297 |
498-627 |
Sentence |
denotes |
Single-distance tomogram recordings with about 1500 projections, and flat images before and after the scan, took less than 2 min. |
T298 |
628-825 |
Sentence |
denotes |
For these scans, the focusing optics (KB-mirrors and waveguide) as well as the fastshutter (Cedrat technologies) were moved out of the beam, and beam size was adjusted by the upstream slit systems. |
T299 |
826-916 |
Sentence |
denotes |
To avoid detector saturation, the beam was attenuated by 4× single crystal silicon wafers. |