Data acquisition and analysis were done using xPONENT software www.luminexcorp.com/xponent/). |
T410 |
94-152 |
Sentence |
denotes |
Data quality was examined based on the following criteria: |
T411 |
153-267 |
Sentence |
denotes |
The standard curve for each analyte has a 5P R2 value > 0.95 with or without minor fitting using xPONENT software. |
T412 |
268-459 |
Sentence |
denotes |
To pass assay technical quality control, the results for two controls in the kit needed to be within the 95% of CI (confidence interval) provided by the vendor for >25 of the tested analytes. |
T413 |
460-535 |
Sentence |
denotes |
No further tests were done on samples with results out of range low (<OOR). |
T414 |
536-710 |
Sentence |
denotes |
Samples with results that were out of range high (>OOR) or greater than the standard curve maximum value (SC max) were not tested at higher dilutions without further request. |