PMC:7120874 / 20964-21493 JSONTXT

Annnotations TAB JSON ListView MergeView

{"target":"https://pubannotation.org/docs/sourcedb/PMC/sourceid/7120874","sourcedb":"PMC","sourceid":"7120874","source_url":"https://www.ncbi.nlm.nih.gov/pmc/7120874","text":"Input spectrum tag (which is the sample name) and the following spot protocols: Starting laser intensity to 165.\nStarting detector sensitivity to 9.\nHighest mass of detection to 200,000 Da.\nOptimal mass range of detection from m/z of 2000 to 20,000 (signals from m/z of 0–2000 are not analyzed as artifacts can be produced by energy-absorbing molecules or other chemical contaminants at this mass range).\nFocus lag time at 800 ns.\nMass deflector to Auto.\nData acquisition method to SELDI quantitation.\n338 laser shots per sample.","divisions":[{"label":"p","span":{"begin":80,"end":112}},{"label":"p","span":{"begin":113,"end":148}},{"label":"p","span":{"begin":149,"end":189}},{"label":"p","span":{"begin":190,"end":404}},{"label":"p","span":{"begin":405,"end":430}},{"label":"p","span":{"begin":431,"end":454}},{"label":"p","span":{"begin":455,"end":501}}],"tracks":[]}