T1 393-580 Epistemic_statement denotes For these reasons it has been used extensively for fever screening at airport quarantine stations in many countries since the outbreak of severe acute respiratory syndrome (SARS) in 2003.
T2 581-669 Epistemic_statement denotes 1, 2 However, many thermography systems used in airports are expensive and non-portable.
T3 1140-1319 Epistemic_statement denotes Second, the thermopile array was designed as a compact module to facilitate its integration into medical applications such that clinicians would find the device convenient to use.
T4 2077-2139 Epistemic_statement denotes However, many of these systems are expensive and non-portable.
T5 2637-2849 Epistemic_statement denotes Results: The maximum facial temperature, measured by the array at 0.3 m from the subject, exhibited a positive correlation with axillary temperature measured using a contact-type thermometer (r = 0.71, p < 0.01).
T6 3063-3321 Epistemic_statement denotes Conclusions: Our cost-effective thermopile array appears promising for future close-range fever screening of patients with infectious diseases at primary care doctor clinics, health care centers, and quarantine stations in developing and developed countries.
T7 3896-4151 Epistemic_statement denotes Because the facial temperature is much higher than the surrounding ambient temperature, the maximum facial temperature (around the inner corner of the eye) can be derived from a two-dimensional thermal facial image by simple image processing (Figure 1b) .
T8 5192-5327 Epistemic_statement denotes The thermopile array measurements exhibited a positive correlation with the contact-type thermometer measurements (r = 0.71, p < 0.01).
T9 5773-5889 Epistemic_statement denotes This approach has been used and generally approved in some recent studies on the evaluation of thermography systems.
T10 6159-6367 Epistemic_statement denotes The explanation for this is that although the thermopile array accurately measures human body temperature, some of the patients were misdiagnosed as normal because they were taking medication to reduce fever.
T11 6368-6495 Epistemic_statement denotes The results indicated that the thermopile array accurately captured elevated human body temperature in close-range thermometry.
T12 6765-6991 Epistemic_statement denotes Commercial thermal imaging systems typically cost several thousand dollars; 6 meanwhile, some researchers have developed inexpensive do-it-yourself infrared scanners to promote the widespread use of thermal imaging technology.
T13 7098-7334 Epistemic_statement denotes However, the main limitation of our thermopile array is that the operational distance between the thermopile array and the subject is 0.5 m, as opposed to approximately 1.0-3.0 m for some high-resolution commercial thermography systems.