Improved specimen adequacy using jumbo biopsy forceps in patients with Barrett's esophagus. AIM: To assess the sampling quality of four different forceps (three large capacity and one jumbo) in patients with Barrett's esophagus. METHODS: This was a prospective, single-blind study. A total of 37 patients with Barrett's esophagus were enrolled. Targeted or random biopsies with all four forceps were obtained from each patient using a diagnostic endoscope during a single endoscopy. The following forceps were tested: A: FB-220K disposable large capacity; B: BI01-D3-23 reusable large capacity; C: GBF-02-23-180 disposable large capacity; and jumbo: disposable Radial Jaw 4 jumbo. The primary outcome measurement was specimen adequacy, defined as a well-oriented biopsy sample 2 mm or greater with the presence of muscularis mucosa. RESULTS: A total of 436 biopsy samples were analyzed. We found a significantly higher proportion of adequate biopsy samples with jumbo forceps (71%) (P < 0.001 vs forceps A: 26%, forceps B: 17%, and forceps C: 18%). Biopsies with jumbo forceps had the largest diameter (median 2.4 mm) (P < 0.001 vs forceps A: 2 mm, forceps B: 1.6 mm, and forceps C: 2mm). There was a trend for higher diagnostic yield per biopsy with jumbo forceps (forceps A: 0.20, forceps B: 0.22, forceps C: 0.27, and jumbo: 0.28). No complications related to specimen sampling were observed with any of the four tested forceps. CONCLUSION: Jumbo biopsy forceps, when used with a diagnostic endoscope, provide more adequate specimens as compared to large-capacity forceps in patients with Barrett's esophagus.