RESULTS On linear regression analysis, postrepair MV gradients were associated with patient age (coefficient = −0.110, standard error = 0.005, P = 0.034), body surface area (coefficient = 0.905, standard error = 0.340, P = 0.008), implanted annuloplasty ring size (coefficient = −0.181, standard error = 0.018, P < 0.001) and the use of Physio I ring (coefficient = 0.414, standard error = 0.122, P = 0.001). On multivariable analysis, postrepair MV gradient was not associated with overall survival [hazard ratio (HR) 1.034, 95% confidence interval (CI) 0.889–1.203; P = 0.66] or freedom from atrial fibrillation (HR 0.849, 95% CI 0.682–1.057; P = 0.14), but did emerge as a risk factor for MV reintervention (HR 1.378, 95% CI 1.033–1.838; P = 0.029). Two out of 11 reinterventions were performed due to MV stenosis and in both patients, high postrepair gradients were seen readily on predischarge echocardiography.