For the evaluation of gastric emptying, a three-dimensional (3D) volume scan is applied [4, 5, 8, 10, 15–17]. In this sequence, the entire stomach is included in the volume acquisition; from the volume data multiple slices with a slice thickness of 0.5 to 1 cm are reconstructed for further evaluation. To evaluate momentary gastric volume, the use of a multi-receive parallel body synergy coil has been advocated at 1.5 T. Sequence parameters include, for example, a turbo field echo (TFE) sequence, TE = 3.5 ms, TR = 10 ms, field of view 450 mm, rectangular field of view 55%, symmetric reduction 50%, flip angle 25°, 256 × 256 pixels, slice thickness 10 mm, with a total scan duration of 25 s, without breath-hold [15].